4. Hydrocarbon analysis
with a VICI pulsed discharge detector (PDD)
 
Courtesy of Scott-Marrin, Inc.
John Marrin, president

 
No ambient air quality analysis is complete without an accurate quantitation of the toxic organic volatile hydrocarbons. The typical gas standard used in calibrating such an analysis is a mixture of paraffins/BTEX in nitrogen. This chromatogram illustrates how the detector’s sensitivity to the air peak is suppressed when argon is added as a dopant. It also shows the PDD’s remarkable sensitivity to a 30 ppb calibration mixture.
 

 
Sample: 30 ppb paraffins/BTEX blend in nitrogen
Detector mode: Argon photoionization
Detector temp: 40°C
Column: 20 m x 0.053 mm DB-1
Column temp: 30°C for 1 min to 110°C at 8°/min
Sample volume: 250 µL, injection valve
Discharge gas: 1.5% Ar (mole) in helium, 30 mL/min
Dopant gas: None
Atten / Range: 2 x 10-1

Additional chromatograms
1 10 ppt SF6 in ultrapure air
2 30 ppb vinyl chloride in nitrogen
3 Haloalkanes mixture in methanol
4.  Hydrocarbon analysis
5 Impurities in food grade CO 2
6 N2 , O2 , and CO2 in acrylic syrup
7 Natural gas liquid (NGL)
8 n-Paraffin wax standard
9 Pesticide analysis
10 Trace impurities in high purity hydrogen
11 Trace impurities in bulk gases
12 Hydrocarbon analysis (comparing FID and PDD)
13 Impurities in hydrocarbon feedstocks
14 PAH residues in soil from a Superfund site
15 PAH residues in an environmental soil sample spike
16 Herbicides in soil samples using EPA Method 8151
 



Home | PRODUCTS | Search | Catalog | Contact
VICI Prices in USD > Aud Plus Import Freight + 10% GST
© 2001 VICI
VICI ® is a registered trademark
of Valco Instruments Co. Inc. and VICI AG